B&K 4193 1/2" Condenser Microphone

This microphone is specially designed for infrasound measurement. When used with adaptor UC0211, the lower limiting frequency of the preamplifier/microphone is reduced to 0.1 Hz.

PLEASE: do not disassemble microphone - adaptor unit!!

Microphone specifications (full specs here) given including UC0211:

  • Open Circuit sensitivity at 250Hz 2 mV / Pa
  • Frequency response 0,07Hz to 20kHz +- 2dB
  • External 200 Vdc polarization required
  • Noisefloor 29dB(A)
  • Upper limit 162dB at 3% distortion

Individual calibrationchart for []:

 


 